Microchannel Plate (MCP) Gate Modules are used to gate an MCP ON by coupling a positive high voltage pulse onto the fixed bias of the MCP output electrode.
If multiple MCPs are used in a detector, the rear MCP is typically gated ON. In the normal OFF state the MCP bias is set below its working voltage. The MCP is turned ON when the high voltage pulse increases the MCP bias by 500 or 1000 volts.
The GM-MCP3 requires external bias voltages for operation. The GM-MCP3-1000, which outputs a 1000V gate pulse, provides the highest level of signal suppression for the most demanding applications.
Contact Photek for customised versions of these Gate Modules.
Specifications
MCP Gating Module
MCP Gating Module
GM-MCP3-500
GM-MCP3-1000
MCP’s
1
2-3
Normal State
OFF
OFF
Output Pulse Frequency
1 kHz
1 kHz
Output Pulse Width
≥ 10 ns
≥ 10 ns
Voltage
Off
On
Bias (max)
0 V + Bias
+500 V + Bias
±9000 V (External)
0 V + Bias
+1000 V + Bias
±9000 V (External)
Supply Voltage
+12 V DC
+12 V DC
Length x Height x Width
166 x 81.1 x 46.1 mm
166 x 81.1 x 46.1 mm
Weight
~ 420 g
~ 420 g
Operating Temperature
0ºC to 70ºC
0ºC to 70ºC
Example Configurations
Related Topic
Application Title
Lorem ipsum dolor sit amet, consec
tetuer adipiscing elit, sed diam
nonummy nibh euismod tincidunt.
Application Title
Lorem ipsum dolor sit amet, consec
tetuer adipiscing elit, sed diam
nonummy nibh euismod tincidunt.
Application Title
Lorem ipsum dolor sit amet, consec
tetuer adipiscing elit, sed diam
nonummy nibh euismod tincidunt.
Application Title
Lorem ipsum dolor sit amet, consec
tetuer adipiscing elit, sed diam
nonummy nibh euismod tincidunt.
Contact Us
26 Castleham Road St Leonards on Sea East Sussex, TN38 9NS United Kingdom